TL1-R
Product Overview & Recommended Operation |
||||||||||||||||||||
The TL1-R thermometer is a compact reference thermometer with accuracy comparable to instrument systems costing twice as much. It has all the features of our standard TL1-A plus a higher level of accuracy. It can be used in applications where extreme accuracy is required. The TL1-R is also perfect for metrology applications and as a calibration reference thermometer. It is an extremely accurate portable hand held instrument that is ideally suited for field applications and indoor laboratory uses. The TL1-R uses a precision, 200 Ohm wire wound platinum resistance temperature sensor and attains its high accuracy through our proprietary calibration algorithm. Standard 5 Point Calibration Report, Nominal Points –0.4, 32, 120, 200, 300 °F. Also available in Higher range sensor and additional points above 300 °F are optional.
|
||||||||||||||||||||
Specifications |
||||||||||||||||||||
|
Product Overview & Recommended Operation
The TL1-R thermometer is a compact reference thermometer with accuracy comparable to instrument systems costing twice as much. It has all the features of our standard TL1-A plus a higher level of accuracy. It can be used in applications where extreme accuracy is required. The TL1-R is also perfect for metrology applications and as a calibration reference thermometer. It is an extremely accurate portable hand held instrument that is ideally suited for field applications and indoor laboratory uses.
The TL1-R uses a precision, 200 Ohm wire wound platinum resistance temperature sensor and attains its high accuracy through our proprietary calibration algorithm.
Standard 5 Point Calibration Report, Nominal Points –0.4, 32, 120, 200, 300 °F.
Higher range sensor and additional points above 300 °F are optional. Also available in °C.

Specifications
Display: |
LCD with 0.25 inch characters Trend Indicating Arrows Low Battery & Fault indication |
User Functions: |
Auto Off after 20 minutes or continuous operation Lowest Reading, Highest Reading, Average Reading Switchable in Units of °F or °C Display hundredths or thousandths resolution |
Construction: |
Stainless Steel temperature sensor Standard Sensor length 8” (20cm) Sensor diameter 1/4” (6.35mm) Sensor mount acetyl plastic Electronics and display in 3.5” x 1.2” anodized Aluminum enclosure Molded Silicone seal Highly durable display window, graphics and membrane switches |
Weight: |
4 ounces, 124 grams |
Standard Range: |
0 to 300°F, -18 to 150°C |
Standard Calibration: |
0 to 300°F, -18 to 150°C Tolerance: +/- 0.07 °F, +/- 0.04 °C (@ 27°C ambient temp.) Resolution: 0.001 NIST Traceable Calibration |
Operating Range: |
32 to 112 °F , 0 to 49°C Recommended Calibration Frequency: 1 yr. |
Temperature Response: |
0 to 50 °C +/- 0.1°C in 3 Minutes, water bath |
Electronics: |
Microcontroller and precision 20 bit A/D Converter FLASH memory Minimum battery temp 4°C Maximum battery temp. 75 °C 200 Ohm, Wire Wound Platinum RTD |
Battery: |
20mm Coin Size 3V Lithium, Type 2032 400 hour service life at room temperatures. |